              Digital Karl Fischer Water Content Tester TP-6AApplication:This TP-6A water content tester adopts the global recognized classical moisture testing method - Karl Fischer Coulometric Method which is the most accurate method with low cost. With the universal standard and analyzing method, it can quickl......

              Send Inquiry

              Product Description

              Digital Karl Fischer Water Content Tester TP-6A
              Application:
              This TP-6A water content tester adopts the global recognized classical moisture testing method - Karl Fischer Coulometric Method which is the most accurate method with low cost. With the universal standard and analyzing method, it can quickly analyze the moisture content in liquid, solid and gas and directly display testing results, with testing range from 0.0001%(1ppm) to 100%. Therefore, this testing method has been widely used in institutions and enterprises of petroleum, chemical industry, power, pharmaceuticals, pesticide, railways, research, etc.

              The whole testing process is automatically controlled by this tester itself. Its auto-stirring and analyzing would be finished within 60seconds, and results would be displayed directly. Totally closed titration system avoids damage to operator from reagent. Equipped with 500ml Karl Fischer Reagent, which could test about 1000 times sample continuously, completely avoiding big deviations of other moisture testing method and thus improving quality index of products.  
                          
              Conform to Standards as follow
              GB6283-86 Chemical Products-Determination of Water Karl Fischer Method(General Method)
              GB/T606-2003 Chemical Reagent-General Method of Determination of Water(Karl Fischer Method)
              GB/T11133-1989 Liquid Petroleum Products-Determination of Water(Karl Fischer Method)
              SH/T0246 Light-fuel Petroleum Products-Determination of Water(Coulmetric Titration Method)
              GB/T7380-1998 Surface Active Agent-Determination of Water(Karl Fischer Method)
              GB/10670-1989 Industrial Fluomethane-Determination of Trace Water(Karl Fischer Method)
              GB/T3776.1-1983 Pesticide Emulsifier-Determination of Water
              GB/T6023-1999 Industrial Butadiene-Determination of Trace Water(Karl Fischer Coulometric Method)
              GB/T1600, GB7600, B/T 18619.1, SH/T 0255 etc.

              Application Area:
              Liquid Chemical Industry:alcohols, ethers, esters, alkenes, benzenes, amines, organic solvents, phenols and other organic products applicable to Karl Fischer Method for determination
              Petroleum and Power Industry:Insulation Oil, Transformer Oil, Turbine Oil, etc.
              Pharmaceutical Industry :medicine materials
              Pesticide Field:emulsifier
              Other Industry:electrolyte of lithium battery
              Solid Various mineral salt, citric acid, explosives, paraffin and other solids with good solubility
              Gas Natural gas, liquefied gas, Freon, butadiene, chloromethane, etc. 


              Features:
              1. Advanced MPU auto-control system and direct A/D conversion of testing electrode data
              2. Completely closed titration system avoids direct touch between people and the reagent as well as influence of environmental humidity.
              3. Auto-control and compensation of electrical current via MCU contributes to quick balance of reagent.
              4. Highlighted six-digit Nixie tube display, with long service lifetime, high luminance and 0.1 microgram' sensitivity.
              5. Stirring speed is adjustable and steerable via buttons; breakdown can be self-checked via buttons.
              6. Electrolysis and testing indicator lights indicates testing status.
              7. 400mA electrolysis current is automatically controlled, which greatly improves its sensitivity and accuracy.
              8. Sample auto-testing and direct display of results.
               
              Technical Parameters
              Method of displa:Highlighted six-digit Nixie tube display
              Method of Testing:Karl Fischer Coulometry Method
              Testing Range:1ugH2O-100mgH2O
              Moisture Content Range:0.0001%(1ppm)-100%
              Senstivity:0.1ugH2O
              End-point Indication:Voice Warning/end-point light indication
              Accuracy:3ugH20-1mgH2O, deviation: ≤0.3% of indications
              More than 1mgH2O, deviation: ≤0.5% of indications(exclude deviation of sample injection)
              Electrolysis Current:Auto-control within 400mA
              Self-checking:Breakdown self-checking button
              Drift Compensation:Auto-control via MPU
              Stirring Speed:Steerable
              Dimension:350×250×160mm
              Working Environment:Temperature: 5ºC-40ºC
              Power Consumption:<40W
              Power Supply:AC220V±10% 50HZ±2.5HZ
              Weight:About 6kg

              Product photograph
              High Accuracy Digital Karl Fischer Water Content Tester Tp-6A

              High Accuracy Digital Karl Fischer Water Content Tester Tp-6A
              Packing and delivery
              High Accuracy Digital Karl Fischer Water Content Tester Tp-6A

              High Accuracy Digital Karl Fischer Water Content Tester Tp-6A

              High Accuracy Digital Karl Fischer Water Content Tester Tp-6A
              Contact info.

               

              Related Category

              Send Inquiry

              Please Feel free to give your inquiry in the form below. We will reply you in 24 hours.
              Click Refresh verification code

              Leave Your Message

              • Click Refresh verification code